Method Development and Stress Degradation Profile of Umifenovir by UV Spectrophotometry

Abdulrahaman Horan , Dr. R. B. Nalanda *, G. Ramya Priya 

Department of Pharmaceutical Analysis, GITAM School of Pharmacy, GITAM (Deemed to be) University, Visakhapatnam, Andhra Pradesh, India. 

Article Info:

_________________________________________________

Article History:

Received 24 June 2024  

Reviewed 10 August 2024  

Accepted 31 August 2024  

Published 15 Sep 2024  

_________________________________________________

Cite this article as: 

Horan A, Nalanda RB, Priya GR, Method Development and Stress Degradation Profile of Umifenovir by UV Spectrophotometry, Journal of Drug Delivery and Therapeutics. 2024; 14(9):74-82

DOI: http://dx.doi.org/10.22270/jddt.v14i9.6789     _________________________________________________

*Address for Correspondence:  

Dr. R. B. Nalanda, Department of Pharmaceutical Analysis, School of Pharmacy, GITAM University, Visakhapatnam, A.P., India-530045. 

Abstract

____________________________________________________________________________________________________________

A simple, selective, and economical UV spectrophotometric method has been developed using ethanol as solvent (50%) to determine the umifenovir content in bulk and pharmaceutical dosage formulations. At a pre-determined λ max of 260 nm, it has shown linear in the 10-50 µg/mL range and exhibited a good correlation coefficient (R2 = 0.9927) and excellent mean recovery (98.00–104%). The method was validated statistically, and recovery studies were done for linearity, precision, and accuracy. The obtained results proved that the method can be employed for the routine analysis of umifenovir in bulks and commercial formulations.

Keywords: UV spectrophotometry, Umifenovir, Validation, Method development, Stress Degradation

 


 

INTRODUCTION 

The chemical name for Umifenovir is ethyl 6-bromo4-[(dimethylamino)methyl]-5-hydroxy-1-methyl-2- (phenylsulfanylmethyl)indole-3-carboxylate. It has a molecular formula of C22H28BrClN2O4S and a molecular weight of 513.9 g/mol. Arbidol, or Umifenovir, is an antiviral medication used in Russia and China to treat influenza and COVID infections.

The chemical structure of broad-spectrum Arbidol is given below: 

 image

Figure 1: Chemical structure of Umifenovir.

Umifenovir is an indole-based, hydrophobic, dual-acting direct antiviral/host-targeting agent used for treating and prophylaxis influenza and other respiratory infections.

Umifenovir is currently licensed in China and Russia for the prophylaxis and treatment of influenza and other respiratory viral infections.

Umifenovir (arbidol) has been investigated in the past as a potential drug for SARS and MERS. Its mechanism of action is similar to Imatinib, an Abelson kinase inhibitor (Abl), the anchor drug in treating Chronic Myeloid Leukaemia. Both molecules prevent the virus from binding to the cell membrane. A trial on 33 adults with laboratory proven COVID-19, who had not been invasive, has reached encouraging favorable results 4- joint therapy of umifenovir and lopinavir/ritonavir was more efficacious than lopinavir/ritonavir only 1. Patients treated not only with protease inhibitor but also with umifenovir became sooner SARS-CoV-19- negative (nasopharyngeal specimens), and more of them were found to improve radiologically, according to CT scans. As reported, umifenovir might decrease both the risk of SARS-CoV-19 transmission and the risk of acute respiratory distress syndrome (ARDS)2. Other studies on the effectiveness of umifenovir showed its superiority in comparison with lopinavir/ritonavir, potency to reduce COVID-19 symptoms and accelerate recovery time.

The literature survey revealed that a few analytical methods, such as HPLC 3, 4, UPLC 5, HPTLC 6, and LC-MS 7, procedures for determining Arbidol have been reported. Hence, the present work aims to develop a simple, precise, and accurate method for estimating Arbidol Hydrochloride in pharmaceutical dosage form using the UV method and validate the developed method per ICH guidelines.

MATERIALS AND METHODS

Instrumentation

A double beam UV-VIS spectrophotometer (UV-1800, Shimadzu) loaded with spectra manager software UV Probe was employed with a spectral bandwidth of 1 nm and wavelength accuracy of ± 0.3 nm with a pair of 10 mm matched quartz cells. The 200-400 nm wavelength range was selected for scanning with medium scanning speed.

Chemicals and reagents

Analytical grade solvents and chemicals used included methanol (CH3OH), disodium hydrogen phosphate (Na2HPO4), potassium dihydrogen phosphate (KH2PO4), and distilled water. Umifenovir was obtained from our research collaborators.

Selection of solvent 

The solubility of Umifenovir was determined in various solvents as pharmacopeia standard. Solubility test was carried out in different solvent like distilled water, methanol, ethanol and chloroform. Both the actives showed good solubility profile in methanol which is inexpensive and thus it was chosen for the study.

Preparation of phosphate buffer 6.8 pH   

About 2.8g of disodium hydrogen phosphate and 1.145g of potassium dihydrogen phosphate were taken and dissolved in sufficient water to produce 100. 

Preparation of standard stock solution (1000µg/ml)

10 mg of umifenovir was weighed and placed into a 10 mL volumetric flask. Afterward, methanol was added to dissolve umifenovir, obtaining a 1000 µg/mL concentration. Further, the dilutions were made from this solution; the 0.3ml was made up to 10 ml using the phosphate buffer to obtain the final concentration of 30 µg/mL.

Determination of λ max 

The standard solution of Umifenovir (30 μg/ml) was scanned in the wavelength region of 200 to 400 nm, and the spectrum was recorded. Phosphate buffer (pH 6.8) was used as blank. It was observed that the maximum wavelength was to be 260 nm by plotting a spectrum between absorbance and wavelength.


 

 

image

Figure 2: Spectrum of Umifenovir at 30 μg/mL

 


 

METHOD VALIDATION

The method developed was according to International Council of Harmonization (ICH) Q2(R1) guidelines. The determined parameters are linearity, accuracy, stress degradation, quantification limit, and detection limit 8, 9.

Linearity

The linearity was assessed by analyzing standard solutions having concentrations (10 -50µg/mL). A plot was constructed for a concentration between absorbance. Standard solutions of umifenovir at different concentrations were prepared. Calibration curves were constructed by plotting the absorbance vs concentration.

Precision

 It is usually expressed as the closeness of agreement between a series of measurements obtained. It is generally expressed as the relative standard deviation of a series of measurements. Intra-day and inter-day studies demonstrated the precision of the method. In the intra-day study, three solutions of the same concentrations (30 µg/mL) were prepared and analyzed thrice daily (morning, afternoon, and evening). In the inter-day variation study, the solutions of the same concentration (30µg/ml) were prepared and analyzed daily for three consecutive days, and the absorbance was recorded. The precision, intra-day and inter-day study results are shown in Tables 3 and 4.

Accuracy

This study was carried out using pre-formulated granules containing pure umifenovir and common excipients. Calculation was done from the label claim and the average weight of the final product. Dilution pattern was followed for the granules to obtain three concentrations—50%, 100%, and 150% of reference solution.

 Limit of Detection (LOD)

Limit of Detection (LOD) is the lowest possible concentration at which the method can detect the analyte within the matrix . It is also defined as the lowest concentration that can be separated from a background noise with some reliability. 

It was found to be 0.5864 µg/ml which is calculated from the following formula, 

      LOD=3.3 σ/S 

Where,

        σ=Standard deviation of the response of the analyte, 

          S=Slope of the linearity plot of the analyte.

Limit of Quantification (LOQ)

Quantitation limit is the concentration that can be quantitated reliably with a specified level of accuracy and precision. It was found to be 1.7769. Limit of Quantitation can be calculated from the following formula,

        LOQ =10σ/S

 

 Where, 

σ=Standard deviation of the response, 

S=Slope of the linearity plot of the analyte. 

STRESS DEGRADATION STUDIES

The ICH guideline Q1A (R2) emphasizes the importance of stress testing for an established analytical method. The stress testing of a drug molecule helps to interpret the API's innate stability properties and validate the analytical process's stability-indicating power. In the current study, stress degradation studies on active samples of moclobemide were performed under several stress environments like acidic, alkali, oxidative, thermal, and photolytic conditions.

Acid Hydrolysis

Acid degradation was performed by taking 1 mL of the stock solution and treating it with 1 mL of 0.1 N HCL. The volume was made up to 10 mL of methanol and stored at room temperature for 3 hours. Aliquots(1.0mL) of the sample were drawn after 60 min, neutralized with NaOH, and diluted appropriately to get a solution of 10µg/mL and analyzed at 260 nm against a blank (solvent without compound). The samples were scanned in the UV region.

Alkaline stress

The procedure described above for the acid hydrolysis was used here for studying alkaline stress in 0.1N NaOH solution.

Pipette  1   mL  of  sample   solution  from  the   mixed 

solution into a 10 mL volumetric flask with precision. 

For 2 hours, it w

Pipette  1   mL  of  sample   solution  from  the   mixed 

solution into a 10 mL volumetric flask with precision. 

For 2 hours, it w

Thermal stress

The Umifenovir sample was taken in a Petri plate and exposed to thermal energy in a hot air oven at a temperature of 45°C for 30 minutes. After 48 hours, 10 mg of the sample was diluted with 10 mL of methanol. From this solution, dilutions were carried out to achieve the appropriate concentration (30 μg/mL), the solution was taken in the cuvette for the UV analysis, and the degradation percentage was calculated.

Photolytic stress

The compound was subjected to various photolytic conditions. Briefly, the sample (1.0mg/µg/mL) was exposed to UV irradiation for 2 hours under a UV lamp at 365 nm. Pure methanol was also exposed to the same conditions. The sample solution was diluted appropriately to get the final 30 µg/mL concentration and analyzed at 260 nm. The solution was also scanned in the UV region to compare the spectrum with the unexposed compounds.

 

 

RESULTS AND DISCUSSION

Method development and optimization

Umifenovir (hydrochloride) is soluble in organic solvents such as ethanol, DMSO, and dimethylformamide. It is sparingly soluble in aqueous buffers. Identifying the wavelength of maximum absorbance is a prerequisite for quantitative UV analysis. A solution representing an absorbance value less than one is generally considered suitable for determining the wavelength of maximum absorbance. Considering the prerequisite and the suitability, the maximum wavelength for the Umifenovir solution (30 µg/mL) was determined using a UV-visible spectrophotometer's full scan mode. The full scan was processed using UV software, and the λmax was identified with the help of software. It was found to be 260 nm for Umifenovir.

METHOD VALIDATION

Linearity

The calibration curve (figure 2) obtained was evaluated by its correlation coefficient. The absorbance of the samples in the range of 10-50 µg/mL (Table 1) was linear with a correlation coefficient (R2) of 0.9927. The LOD and LOQ were calculated as 0.5864 mg/mL and 1.7769 mg/mL respectively.

Precision

The developed method's intra-day and inter-day precision study (Tables 3 and 4) confirmed adequate sample stability and method reliability where all the RSDs were <2%.

Accuracy

 Results within the range of 99.00–100.07% ensure an accurate method (Table 5) and indicate non-interference with the excipients of the formulation.

Table 1: Linearity for Umifenovir

Concentration (µg/mL)

Absorbance

10

0.2951

15

0.3413

20

0.5253

25

0.7425

30

0.9189

35

1.0583

40

1.2121

45

1.4010

50

1.4995

          

image

Figure 2: Calibration curve of Umifenovir


 

 

 

 

Figure 3: Overlay spectrum of Umifenovir at 10-50 μg/mL

Table 2: Optical characteristics and other parameters:

Parameters

Method

Reagent

Phosphate buffer pH 6.8

Linearity (µg/mL)

10-50

Max(nm)

260

Molar extinction (litre/mole/cm)

1.455x10 3

Sandell’s sensitivity (µg/cm2/0,001 absorbance unit)

0.0326

Slope

0.0323

Intercept

0.0804

Correlation coefficient

0.9927

Precision (% RSD)       Intra-day

0.86 -1.3

0.86 – 1.7

                                           Inter-day

Accuracy(%Recovery)

98.61-104.37

 

Table 3: Intraday precision


 

Morning                               

Conc (µg/mL)

Absorbance

Obs. conc

Accuracy%

30

0.933

31.37

104.58

30

0.921

30

103.34

30

0.914

30.78

102.62

30

0.919

30.94

103.13

30

0.923

31.06

103.55

30

0.934

31.40

104.68

MEAN

0.924

 

STD. DEV

0.007

 

%RSD

0.8603

 

 

Afternoon

Conc (µg/mL)

Absorbance

Obs. conc

Accuracy%

30

0.912

30.27

102.41

30

0.932

31.34

104.47

30

0.927

31.18

103.96

30

0.906

30.53

101.79

30

0.902

30.41

101.38

30

0.924

31.09

103.65

MEAN

0.9171

 

STD. DEV

0.0122

 

%RSD

1.3307

 


 

Evening

Conc (µg/mL)

Absorbance

Obs.conc

Accuracy%

30

0.933

31.37

104.58

30

0.926

31.15

103.85

30

0.922

31.03

103.44

30

0.916

30.84

102.82

30

0.908

30.60

102.00

30

0.911

30.69

102.31

MEAN

0.9193

 

STD. DEV

0.009

 

%RSD

1.0288

 

 

 

 Table 4: Inter day precision 

 Day – 1 

S. No

Conc (µg/mL)

Absorbance

Obs.conc

Accuracy%

1

30

0.931

31.31

104.37

2

30

0.926

31.15

103.85

3

30

0.927

31.18

103.96

4

30

0.917

30.87

102.93

5

30

0.922

31.03

103.44

6

30

0.909

30.63

102.10

MEAN

 

0.922

 

 

STD.DEV

 

0.008

 

 

%RSD

 

0.862

 

 

 

Day – 2

S. No

Conc (µg/mL)

Absorbance

Obs.conc

Accuracy%

1

30

0.932

31.34

104.47

2

30

0.933

31.37

104.58

3

30

0.911

30.69

102.31

4

30

0.932

31.34

104.47

5

30

0.916

30.84

102.82

6

30

0.907

30.56

101.89

MEAN

 

0.922

 

 

STD.DEV

 

0.012

 

 

%RSD

 

1.286

 

 

 

Day – 3

S. No

Conc (µg/mL)

Absorbance

Obs.conc

Accuracy%

1

30

0.935

31.43

104.78

2

30

0.9

30.35

101.78

3

30

0.921

31.00

103.34

4

30

0.946

31.77

105.92

5

30

0.929

31.25

104.16

6

30

0.918

30.91

103.03

MEAN

 

0.925

 

 

STD.DEV

 

0.016

 

 

%RSD

 

1.707

 

 

 

 

 

 

 

Table 5: Accuracy for Umifenovir

Level

Added amount (µg/mL)

Absorbance

Amount found(µg/mL)

%Recovery

Mean recovery

S.D.

%RSD

 

50

 

15

0.415

15.33

102.24

 

101.69

 

0.630

 

 

0.620

 

15

0.409

15.15

101.01

15

0.413

15.27

101.83

 

100

 

30

0.931

31.31

104.37

 

103.44

 

 

0.881

 

 

0.852

 

30

0.914

30.78

102.62

30

0.921

31.00

103.34

 

150

 

45

1.395

45.67

101.50

 

99.39

 

 

1.847

 

 

1.859

 

45

1.345

44.13

98.06

45

1.353

44.37

98.61

 

Table 6: Stress degradation studies

Stress Condition

Reagent

Time(min)

%Assay

%Degradation

Acid Hydrolysis

0.1N HCL

180

95.60

4.39

Alkaline

0.1N NaOH

180

77.18

22.81

Oxidative

1% H2O2

60

77.97

22.02

Thermal

45⁰C

30

93.54

6.45

Photolytic

365nm

120

86.04

13.95

 

image

Figure 4(a): Acid Stress degradation studies

 

image

Figure 4(b): Alkaline Stress degradation studies

image

Figure 4(c): Oxidative Stress degradation studies

 


 

CONCLUSION

The proposed UV- Spectrophotometric method for estimation of Umifenovir in API and its validation was carried out per ICH guidelines. By studying various parameters, we finally conclude that the process is simple, precise, sensitive, economical, and specific and can be applied to determine Umifenovir in quality control laboratories. All the required validation parameters were estimated and were found to be within the limits.

 

Acknowledgments

The authors thank the GITAM Institute of Pharmacy, Visakhapatnam, for providing the necessary research facilities.

Authors contribution

RBN designed the study concept and corrected the manuscript. AH assisted in the data collection, analysis, and interpretation of results and manuscript preparation. RP helped carry out the analysis and manuscript preparation.

Funding source

There is no financial support available for this research

Competing interests/Conflicts of interest

All the authors declared that they have no conflicts of interest

References

1. Dong H, He Y, Ting W, Huan Y, Rong Y and Zongan L. Efficacy and safety of umifenovir for coronavirus disease 2019 (COVID-19): A Systematic review and meta-analysis. Journal of Medical Virology, 2021;93(1):481-490. https://doi.org/10.1002/jmv.26256 PMid:32617989 PMCid:PMC7361300

2. Philippe-Henri S, Hassane SY, Audrey S, Marie A, Lionel T, Olivier T, Victorie V, Najet Y, Bernard D. Intrinsic stability of the antiviral drug umifenovir by stress testing and DFT. Journal of Pharmaceutical and Biomedical Analysis, 2021;196:113934. https://doi.org/10.1016/j.jpba.2021.113934 PMid:33549876

3. Angirekula N, Mathrusri Annapurna M. Quantification of arbidol by RP-HPLC with photodiode array detection. Asian Journal of Pharmaceutics 2018, 12(2):553-557.

4. Srinivasa Rao S, Neelu Jain. Validated stability indicating method for determination of umifenovir-remdesivir in presence of its degradation products. International Journal of Development Research. 2021;11(4):46227-46232. https://doi.org/10.37118/ijdr.21666.04.2021

5. Muzaffar I, Faisal I, Essam AA, Faisal I, Mohd Abul K, Sulaiman SA. A Validated UPLC-MS/MS method for rapidly quantifying Umifenovir in plasma samples and its greenness assessment. Separations. 2023;10(7):379. https://doi.org/10.3390/separations10070379

6. Damle MC, Phadtare SA, Raskar HD, Gadge KG, Mehendre R, Bothara KG. A validated HPTLC method for determination of arbidol from pharmaceutical formulation. International Journal of ChemTech Research. 2010;2(2):1042-1046

7. Xiao L, Yu-Wei H, Juan L, Xiao L, Yu-Wei H, Juan L, Xiao-Bo L, Kai-Shun B. Determination of arbidol in human plasma by LC-ESI-MS. Journal of Pharmaceutical and Biomedical Analysis. 2007;43(1):371-375. https://doi.org/10.1016/j.jpba.2006.06.039 PMid:16884882

8. Validation of analytical procedures: text and methodology, in: International Conference on Harmonization (ICH), Q2(R1), IFPMA, Geneva, Switzerland, 2005.

9. ICH Q1A (R2), International Conference on Harmonization (ICH). Stability Testing of New Drug Substances and Products, IFPMA, Geneva (2003).