Stability indicating analytical method validation for hydralazine hydrochloride related substances method-I by Reverse Phase High Performance Liquid Chromatography in drug substances

  • Swamy K. Reddy Sri Satya Sai University of Technology & Medical Sciences, Sehore (M.P), India-466001
  • Pushpendra Sharma Sri Satya Sai University of Technology & Medical Sciences, Sehore (M.P), India-466001
  • Asheesh Singh Faculty of pharmacy, VNS institute, Neelbud Bhopal (M.P.), India-462044

Abstract

A simple, rapid, precise, accurate and cost effective stability-indicating reversed phase (RP) HPLC related substance method-1 was validated for Hydralazine Hydrochloride (HYD HCl) in Active pharmaceutical ingredient. All the analytical parameters were determined as per ICH Q2B guidelines. Good chromatographic separation was achieved with Inertsil ODS 3V column (4.6 mm x 250 mm, 5 μm particle size) at a wavelength of 230 nm using phosphate buffer pH 2.5 and acetonitrile as mobile phase A and Methanol as mobile phase B with gradient programming with a flow rate of 1.0 ml/ min. The Resolution between Hydralazine peak and impurity-A should not be less than 3.0. From the statistical treatment of the linearity data of Hydralazine HCl, it is clear that the response of Hydralazine HCl is linear between 50 % to 150 % level. The correlation coefficient is greater than 0.998. The developed method showed good linearity, Accuracy, reproducibility, precision and robustness and can be suitably applied for the routine quality control analysis in the estimation of commercial formulations.


Keywords: Hydralazine hydrochloride, HPLC, Validation, Estimation

Downloads

Download data is not yet available.

Author Biographies

Swamy K. Reddy, Sri Satya Sai University of Technology & Medical Sciences, Sehore (M.P), India-466001

Sri Satya Sai University of Technology & Medical Sciences, Sehore (M.P), India-466001

Pushpendra Sharma, Sri Satya Sai University of Technology & Medical Sciences, Sehore (M.P), India-466001

Sri Satya Sai University of Technology & Medical Sciences, Sehore (M.P), India-466001

Asheesh Singh, Faculty of pharmacy, VNS institute, Neelbud Bhopal (M.P.), India-462044

Faculty of pharmacy, VNS institute, Neelbud Bhopal (M.P.), India-462044

References

1. Srinivasa RM, Shanmukha KJV, Ramachandran D, Spectrophotometric estimation of hydralazine hydrochloride in pharmaceutical formulations, Journal of Pharmaceutical Biology, 2012; 2(1):16-19.
2. Mopper B, UV spectrophotometric determination of hydralazine hydrochloride in tablets: collaborative study, Journal - Association of Official Analytical Chemists, 1988; 71(6):1121-1122.
3. Naik DV, Davis BR, Minnet KM, Schulman SG, Fluorescence of hydralazine in concentrated sulfuric acid, Journal of Pharmaceutical Science, 1976: 65:270-274.
4. Gaidukevich, OM, Zhukova TV, Zarechensky MA, Sim G, Sensitive spectrophotometric methods for quantitative determination of hydralazine hydrochloride in pure and pharmaceutical formulation, Khimiko Farmatsevticheskii Zhurnal, 1988; 43:41-49.
5. Adegoke OA, Nwoke CE, Spectrophotometric determination of hydralazine using p dimethyl aminobenzaldehyde, Journal of the Iranian Chemical Society, 2008; 5(2):316-323.
6. International Conference on Harmonization, ICH harmonized tripartite guideline validation of analytical procedures: text and methodology Q2 (R1) ICH, Geneva, 2005.
7. Gelber L, Papas AN, Validation of high-performance liquid chromatographic methods for analysis of sustained release preparations containing nitroglycerin, isosorbide dinitrate, or pentaerythritol tetranitrate. Journal of Pharmaceutical Science, 2013; 72:124-26.
8. Massoud M, Hamidreza F, Babak G, Ladan T, Determination of Isosorbide Dinitrate in serum by Gas Chromatography with new generation of electron captures detector and its application in pharmacokinetic study, Iranian Journal of Pharmaceutical Research, 2008; 8:251-255.
9. Prue DG, Johnson RN, Kho BT, Gas-liquid chromatographic Determination of isosorbide dinitrate in tablets. Journal - Association of Official Analytical Chemists, 2017; 60:1341-44.
10. Yan X, Zhou H, Zhang Z, He D, He C, Determination of hydralazine with flow injection chemiluminescence sensor using molecularly imprinted polymer as recognition element. Journal of pharmaceutical and biomedical analysis, 2006; 41:694-700.
Statistics
291 Views | 264 Downloads
How to Cite
1.
Reddy SK, Sharma P, Singh A. Stability indicating analytical method validation for hydralazine hydrochloride related substances method-I by Reverse Phase High Performance Liquid Chromatography in drug substances. JDDT [Internet]. 15Dec.2018 [cited 18Jan.2021];8(6-s):125-34. Available from: http://jddtonline.info/index.php/jddt/article/view/2099